A non-EU manufacturer enters the EU by classifying the device under MDR or IVDR Annex VIII, appointing an Authorised Representative and a Person Responsible for Regulatory Compliance, registering as an actor in EUDAMED to obtain an SRN, building technical documentation to Annex II and III with a clinical or performance evaluation, undergoing notified-body conformity assessment unless the device is Class I (non-sterile, non-measuring, non-reusable surgical) or IVD Class A (non-sterile), registering each device in EUDAMED before placing it on the market, and then running PMS, PSUR and vigilance on the cadence the class requires.
Classification and routes
The MDR classifies devices into Class I (with sub-categories for sterile, measuring and reusable surgical devices), IIa, IIb and III under the 22 rules of Annex VIII; the IVDR classifies IVDs into Classes A to D under its own Annex VIII. The class decides the conformity-assessment route and therefore how much of the file a notified body reviews. Only Class I devices that are not sterile, measuring or reusable surgical, and IVD Class A devices that are not sterile, self-certify; everything else involves a notified body, from a quality-system audit with technical-documentation sampling to a full design examination. Classification is the first decision in any EU plan and the one most often got wrong by a single sentence of intended purpose.
Technical documentation and evidence
Annex II and III set out the technical documentation: device description and specification, information supplied by the manufacturer, design and manufacturing information, the general safety and performance requirements with the method and evidence for each, benefit-risk and risk management, verification and validation (including biocompatibility, electrical safety, software, usability, stability and the clinical evaluation or performance evaluation), and the post-market documentation. The clinical evaluation under Articles 61 and 62 and Annex XIV, and the performance evaluation under IVDR Articles 56 and 57 and Annex XIII, are living documents fed by PMCF or PMPF. See technical documentation for how the file is built and traced.
Actors, representation and EUDAMED
A non-EU manufacturer appoints an Authorised Representative in the EU under Article 11; the mandate is a legal instrument and the representative carries obligations and potential liability. Every manufacturer has a Person Responsible for Regulatory Compliance under Article 15. Importers and distributors have verification duties under Articles 13 and 14. The manufacturer, the representative and the importer each register in EUDAMED's Actor module and receive a Single Registration Number; devices are registered with their Basic UDI-DI and UDI-DIs in the UDI/Device module before being placed on the market. Since 28 May 2026 these registrations are mandatory and national registers no longer satisfy them; devices already on the market before that date had until 28 November 2026 to be registered.
PMS, PSUR and vigilance
Class I devices need a PMS report under Article 85; Class IIa devices a PSUR under Article 86 at least every two years; Class IIb and III a PSUR at least annually, with IIb implantable and Class III PSURs going to the notified body. IVD Class C and D devices need a PSUR under IVDR Article 81. Serious incidents and field safety corrective actions are reported under Articles 87 to 89 within the statutory timelines, through national routes until EUDAMED's Vigilance module is mandated. Trend reporting under Article 88 needs a method and thresholds. See post-market surveillance.
The targeted revision and the AI Act
Two legislative developments sit over every EU plan in 2026. The Commission's proposal for a targeted revision of the MDR and IVDR (COM(2025) 1023, procedure 2025/0404(COD)) was published on 16 December 2025; Parliament's rapporteur issued a draft report on 1 July 2026, and adoption is targeted for 2027. It is a proposal: the current text applies until an amending regulation is adopted and in force, and a file built to the current requirements will be well placed under any likely outcome.
The AI Act, as amended by the Digital Omnibus on AI (Regulation (EU) 2026/1744, in force 27 July 2026), classifies AI that is a device or a safety component of one requiring notified-body assessment as a high-risk AI system under Article 6(1), with obligations applying from 2 August 2028; Annex III standalone high-risk systems apply from 2 December 2027. The classification route determines the date, and the software pillar states both.
Sequence of steps
- Classify the device under Annex VIII and document the rule applied.
- Appoint the Authorised Representative and the PRRC; register as an actor in EUDAMED and obtain the SRN.
- Select the notified body and open the structured dialogue.
- Build the technical documentation to Annex II and III, including the clinical or performance evaluation and the PMS plan.
- Bring the quality system to ISO 13485:2016 and the MDR's Article 10 requirements.
- Undergo conformity assessment; receive the certificate; draw up the EU declaration of conformity; affix the CE mark.
- Register devices in EUDAMED before placing on the market; assign Basic UDI-DI and UDI-DIs.
- Run PMS, PSUR, PMCF or PMPF and vigilance on the cadence the class requires; keep the clinical or performance evaluation current.
Dated changes to know about
| Change | Date | Status | What it means for you |
|---|---|---|---|
| EUDAMED Actor, UDI/Device, Notified Bodies and Certificates, and Market Surveillance modules become mandatory | 28 May 2026 | In force | New devices must be registered in EUDAMED before placing on the market; actors need an SRN; national registers no longer satisfy these obligations. Vigilance and clinical investigation modules are not yet mandatory. |
| Deadline to register devices placed on the market before 28 May 2026 and still supplied | 28 November 2026 | In force | Legacy and Regulation devices already on the market must be in the UDI/Device module by this date under the Commission's transition timeline. Certificates issued before 28 May 2026 must be registered by notified bodies by 28 May 2027. |
| swissdamed device registration mandatory in Switzerland | 1 July 2026, transition to 31 December 2026 | In force | Devices placed on the Swiss market must be registered in swissdamed; devices subject to vigilance actions have no transition period. |
| Digital Omnibus on AI, Regulation (EU) 2026/1744, amends the AI Act | In force 27 July 2026 | In force | High-risk obligations for AI that is a device or safety component requiring notified-body assessment (Article 6(1)) apply from 2 August 2028; Annex III standalone systems from 2 December 2027. |
| Targeted revision of the MDR and IVDR, Commission proposal COM(2025) 1023 | Proposed 16 December 2025; Parliament draft report 1 July 2026; adoption targeted 2027 | Proposed | Negotiation continues; the current MDR and IVDR text applies in full until an amending regulation is adopted and in force. |
Legal basis and references
| Reference | Edition or version cited | Where it applies |
|---|---|---|
| Regulation (EU) 2017/745 (MDR) | As amended by Regulations (EU) 2020/561, 2023/607 and 2024/1860 | Medical devices — classification, conformity assessment, obligations of economic operators, PMS and vigilance |
| Regulation (EU) 2017/746 (IVDR) | As amended by Regulations (EU) 2022/112, 2023/607 and 2024/1860 | In vitro diagnostic medical devices |
| Commission Decision (EU) 2025/2371 | Published 27 November 2025 | Functionality of the first four EUDAMED modules; mandatory use from 28 May 2026 |
| MDCG 2021-24; MDCG 2019-11; MDCG 2020-1; MDCG 2022-2; MDCG 2022-21; MDCG 2023-3; MDCG 2019-16 | Current revisions | Classification; software; clinical and performance evaluation; PSUR; vigilance; cybersecurity |
| Harmonised standards published in the Official Journal under the MDR and IVDR | As listed at the review date | Presumption of conformity; harmonisation status is distinct from ISO publication date |
| Regulation (EU) 2024/1689 (AI Act) as amended by Regulation (EU) 2026/1744 | In force 27 July 2026 | AI-enabled devices |
| Swiss Medical Devices Ordinance (MedDO) and IvDO | As amended; swissdamed registration mandatory from 1 July 2026 | Switzerland |
Boundary
Authentcare prepares the technical documentation and coordinates the notified-body relationship; the notified body certifies, and the competent authorities and Commission regulate. EUDAMED registrations are made in the manufacturer's or Authorised Representative's own account.